Reg. No. 200 Accreditation for testing and calibration | Page: 1
2
3
4
5
6
7
all
|
EA Scope |
Equipment |
Quantity |
Range Low Limit |
Range High Limit |
U(CMC) Low Limit |
U(CMC) High Limit |
Working Standard |
Method |
Remark |
In Situ |
Location |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
0
μA |
200
μA |
0,68
nA |
2,1
nA |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
200
μA |
2
mA |
7,9
nA |
21
nA |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
2
mA |
20
mA |
82
nA |
0,24
μA |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
20
mA |
200
mA |
1,9
μA |
8,7
μA |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
200
mA |
2
A |
35
μA |
0,21
mA |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (10 Hz - 3 kHz
)
|
200
mA |
2
A |
700
ppm |
0,18
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (3 kHz - 30 kHz
)
|
200
mA |
2
A |
700
ppm |
0,31
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (10 Hz - 3 kHz
)
|
2
A |
20
A |
900
ppm |
0,35
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (3 Hz - 10 kHz
)
|
2
A |
20
A |
0,26
% |
0,35
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (10 Hz - 3 kHz
)
|
100
μA |
200
μA |
580
ppm |
680
ppm |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (3 kHz - 30 kHz
)
|
100
μA |
200
μA |
580
ppm |
0,42
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
2
A |
20
A |
0,81
mA |
4,7
mA |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (10 Hz - 3 kHz
)
|
200
μA |
2
mA |
380
ppm |
0,13
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (3 kHz - 30 kHz
)
|
200
μA |
2
mA |
380
ppm |
0,51
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (3 kHz - 30 kHz
)
|
2
mA |
20
mA |
380
ppm |
0,42
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (10 Hz - 3 kHz
)
|
2
mA |
20
mA |
380
ppm |
0,13
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (10 Hz - 3 kHz
)
|
20
mA |
200
mA |
580
ppm |
0,15
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (3 kHz - 30 kHz
)
|
20
mA |
200
mA |
580
ppm |
0,42
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Power (45 Hz - 65 Hz
)
|
82,5
mW |
100
mW |
0,59
% |
0,12
% |
Electrical Calibrator |
Internal procedure: 4330 |
PF = 1 |
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Power (45 Hz - 65 Hz
)
|
0,1
W |
3000
W |
0,12
% |
0,14
% |
Electrical Calibrator |
Internal procedure: 4330 |
PF = 1 |
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Power (45 Hz - 65 Hz
)
|
3
kW |
21
kW |
0,14
% |
0,21
% |
Electrical Calibrator |
Internal procedure: 4330 |
PF = 1 |
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Power (45 Hz - 65 Hz
)
|
82,5
mW |
100
mW |
0,58
% |
0,16
% |
Electrical Calibrator |
Internal procedure: 4330 |
PF < 1; U(CMC) relative to VA |
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Power (45 Hz - 65 Hz
)
|
0,1
W |
3000
W |
0,15
% |
0,15
% |
Electrical Calibrator |
Internal procedure: 4330 |
PF < 1; U(CMC) relative to VA |
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Power (45 Hz - 65 Hz
)
|
3
kW |
21
kW |
0,24
% |
0,24
% |
Electrical Calibrator |
Internal procedure: 4330 |
PF < 1; U(CMC) relative to VA |
|
El-Aarhus |
Time and frequency |
Time and frequence meas. eq. |
Timeinterval |
1
μs |
1
s |
6,6
ns |
7,1
μs |
counter |
Internal procedure: D1-9.3 |
|
|
El-Aarhus |
Time and frequency |
Time and frequence meas. eq. |
Timeinterval |
1
s |
1
ks |
7,1
μs |
7,1
ms |
counter |
Internal procedure: D1-9.3 |
|
|
El-Aarhus |
Time and frequency |
Time and frequence meas. eq. |
Frequency |
100
mHz |
10
MHz |
12
pHz |
1,2
mHz |
GPSDO and counter |
Internal procedure: D1-9.1, D1-9.2 |
|
|
El-Aarhus |
Time and frequency |
Time and frequence meas. eq. |
Frequency |
10
MHz |
225
MHz |
1,2
mHz |
26
mHz |
GPSDO and counter |
Internal procedure: D1-9.1, D1-9.2 |
|
|
El-Aarhus |
Time and frequency |
Time and frequence meas. eq. |
Period |
4,4
ns |
1
μs |
0,00052
fs |
0,12
fs |
GPSDO and counter |
Internal procedure: D1-9.1, D1-9.2 |
|
|
El-Aarhus |
Time and frequency |
Time and frequence meas. eq. |
Period |
1
μs |
10
s |
0,12
fs |
1,2
ns |
GPSDO and counter |
Internal procedure: D1-9.1, D1-9.2 |
|
|
El-Aarhus |
Temperature |
Pyrometers and IR equipments |
Temperature (suface) |
-40
°C |
80
°C |
0,3
°C |
0,3
°C |
Reference thermometer, blackbody |
Intern procedure D1-15 |
|
|
TERM-Aarhus |
Time and frequency |
Time and frequence meas. eq. |
Timeinterval |
30
s |
172800
s |
0,038
s |
0,043
s |
GPS / WaveForm Generator / Counter |
Internal procedure Calibration of stopwatches (ID:9777) |
Time Interval (Calibration of Stopwatches). |
|
El-Aarhus |
Temperature |
Pyrometers and IR equipments |
Temperature (suface) |
80
°C |
270
°C |
0,6
°C |
0,6
°C |
Reference thermometer, blackbody |
Intern procedure D1-15 |
|
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature (air) |
-40
°C |
0
°C |
0,2
°C |
0,13
°C |
Digital thermometer |
Internal procedure D1-6.1 |
Air temperature |
|
FUGT-Aarhus |
Temperature |
Thermometers |
Temperature (air) |
0
°C |
70
°C |
0,028
°C |
0,049
°C |
Digital thermometer |
Internal procedure D1-6.1 |
Air temperature |
|
FUGT-Aarhus |
Temperature |
Thermometers |
Temperature (air) |
70
°C |
90
°C |
0,17
°C |
0,25
°C |
Digital thermometer |
Internal procedure D1-6.1 |
Air temperature |
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Relative humidity of air (5 - 20 °C
)
|
5
%rh |
10
%rh |
0,16
%rh |
0,25
%rh |
Dew point instrument |
Internal procedure D1-14 |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Relative humidity of air (20 - 30 °C
)
|
0,2
%rh |
10
%rh |
0,12
%rh |
0,2
%rh |
Dew point instrument |
Internal procedure D1-14 |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Relative humidity of air (30 - 50 °C
)
|
5
%rh |
10
%rh |
0,14
%rh |
0,17
%rh |
Dew point instrument |
Internal procedure D1-14 |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Relative humidity of air (0 - 70 °C
)
|
10
%rh |
95
%rh |
0,14
%rh |
0,48
%rh |
Dew point instrument |
Internal procedure D1-6.1 |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Relative humidity of air (70 - 80 °C
)
|
10
%rh |
66
%rh |
0,34
%rh |
1,1
%rh |
Dew point instrument |
Internal procedure |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Relative humidity of air (80 - 90 °C
)
|
10
%rh |
45
%rh |
0,34
%rh |
0,73
%rh |
Dew point instrument |
Internal procedure D1-6.1 |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Dewpoint temperature |
-75
°C |
-50
°C |
0,19
°C |
0,056
°C |
Dew point instrument |
Internal procedure D1-14 |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Dewpoint temperature |
-50
°C |
70
°C |
0,056
°C |
0,085
°C |
Dew point instrument |
Internal procedure D1-6.1 and D1-14 |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Dewpoint temperature |
-75
°C |
-50
°C |
0,2
°C |
0,14
°C |
Dewpoint instrument |
Internal procedure D1-14 |
Pressure dewpoint: 0 - 7 bar |
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Dewpoint temperature |
-50
°C |
10
°C |
0,14
°C |
0,14
°C |
Dewpoint instrument |
Internal procedure D1-14 |
Pressure dewpoint: 0 - 7 bar |
|
FUGT-Aarhus |
Humidity |
Moisture meter |
Water content, g(H2O)/g(dry) |
8
% |
8
% |
0,5
% |
0,5
% |
Balance |
Internal procedure: 8220 |
Wood type: spruce, fir not impregnated |
|
FUGT-Aarhus |
Humidity |
Moisture meter |
Water content, g(H2O)/g(dry) |
17
% |
17
% |
0,8
% |
0,8
% |
Balance |
Internal procedure: 8220 |
Wood type: spruce, fir not impregnated |
|
FUGT-Aarhus |
Humidity |
Moisture meter |
Water content, g(H2O)/g(dry) |
20
% |
20
% |
0,9
% |
0,9
% |
Balance |
Internal procedure: 8220 |
Wood type: spruce, fir not impregnated |
|
FUGT-Aarhus |
Humidity |
Moisture meter |
Water content, g(H2O)/g(dry) |
26,7
% |
26,7
% |
1,1
% |
1,1
% |
Balance |
Internal procedure: 8220 |
Wood type: spruce, fir not impregnated |
|
FUGT-Aarhus |
|