Reg. No. 200 Accreditation for testing and calibration |
EA Scope |
Equipment |
Quantity |
Range Low Limit |
Range High Limit |
U(CMC) Low Limit |
U(CMC) High Limit |
Working Standard |
Method |
Remark |
In Situ |
Location |
Flow |
Flowmeters |
Volume flow rate l/s (1 ml/h - 100 ml/h
)
|
0,278
μl/s |
27,8
μl/s |
0,05
% |
4
% |
Weighing machine |
Internal procedure D1-30 |
1....100 ml/h |
|
FLOW-Aarhus |
Flow |
Flowmeters |
Volume flow rate l/s (100 ml/h - 6000 ml/h
)
|
27,8
μl/s |
1,68
ml/s |
0,05
% |
0,05
% |
Weighing machine |
Internal procedure D1-30 |
100....6000 ml/h |
|
FLOW-Aarhus |
Flow |
Flowmeters |
Volume flow rate (5 l/h - 40 m^3/h
)
|
1,4
mL/s |
11
L/s |
0,05
% |
0,1
% |
Weighing machine |
Internal procedure D1-1 D1-8 |
0,005...40m3/h Watertemperature 4...85°C; Calibration and reverification acc. to OIML R49, ISO/EN 4064, EN1434 |
|
FLOW-Aarhus |
Flow |
Flowmeters |
Volume flow rate (40 m^3/h - 500 m^3/h
)
|
11
L/s |
139
L/s |
0,05
% |
0,1
% |
Weighing machine |
Internal procedure D1-1 D1-8 |
40...500m3/h Watertemperature 4...85°C; Calibration and reverification acc. to OIML R49, ISO/EN 4064, EN1434 |
|
FLOW-Aarhus |
Flow |
Massflowmeters |
Mass flow rate (1 g/h - 6000 g/h
)
|
0,278
mg/s |
1,67
g/s |
0,05
% |
4
% |
Weighing machine |
Internal Procedure D1-30 |
1 ....6000 g/h |
|
FLOW-Aarhus |
Flow |
Flowmeters |
Volume flow rate (5 l/h - 1000 m^3/h
)
|
0,001389
L/s |
277,78
L/s |
0,3
% |
0,4
% |
Reference meter |
Internal procedure D1-2; D1-9 |
0,005...1000 m3/h Water temperatures 5...85°C |
|
FLOW-Aarhus |
Flow |
Massflowmeters |
Mass flow rate (5 kg/h - 1000 ton/h
)
|
0,001389
kg/s |
277,78
kg/s |
0,3
% |
0,4
% |
Reference meter |
Internal procedure D1-2; D1-9 |
0,005...1000 ton/h Water temperatures 4...85°C |
|
FLOW-Aarhus |
Flow |
Massflowmeters |
Mass flow rate (5 kg/h - 40.000 kg/h
)
|
0,001389
kg/s |
11
kg/s |
0,05
% |
0,05
% |
Weighing machine |
Internal Procedure D1-1 D1-8 |
5...40.000 kg/h watertemperature 4...85°C |
|
FLOW-Aarhus |
Flow |
Massflowmeters |
Mass flow rate (40.000 kg/h - 500.000 kg/h
)
|
11
kg/s |
139
kg/s |
0,05
% |
0,1
% |
Weighing machine |
Internal Procedure D1-1 D1-8 |
40.000...500.000 kg/h watertemperature 4...85°C |
|
FLOW-Aarhus |
Volume |
Volume meters, liquid |
Volume |
1
L |
60
L |
0,05
% |
0,05
% |
Weighing machine |
Internal procedure D1-1 D1-8 |
1...60 litre Watertemperature 4...85°C; Kalibrering og reverifikation iht. OIML R49, ISO/EN 4064, EN1434 |
|
FLOW-Aarhus |
Volume |
Volume meters, liquid |
Volume |
200
L |
5000
L |
0,05
% |
0,1
% |
Weighing machine |
Internal procedure D1-1 D1-8 |
200...5000 litre Watertemperature 4...85°C;Calibration and reverification acc. to OIML R49, ISO/EN 4064, EN1434 |
|
FLOW-Aarhus |
Volume |
Volume meters, liquid |
Volume in litres |
1
l |
250000
l |
0,4
% |
0,4
% |
Master meter |
Internal procedure D1-2 D1-9 |
1...250000 liter Watertemperature 4...85°C |
|
FLOW-Aarhus |
Flow |
Massflowmeters |
Mass |
1
kg |
60
kg |
0,05
% |
0,05
% |
Weighing machine |
Internal procedure D1-1 D1-8 |
1...60kg Watertemperature 4...85°C; Calibration and reverification acc. to OIML R49, ISO/EN 4064, EN1434 |
|
FLOW-Aarhus |
Flow |
Massflowmeters |
Mass |
200
kg |
5000
kg |
0,05
% |
0,1
% |
Weighing machine |
Internal procedure D1-1 D1-8 |
200...5000kg Watertemperature 4...85°C; Calibration and reverification acc. to OIML R49, ISO/EN 4064, EN1434 |
|
FLOW-Aarhus |
Flow |
Massflowmeters |
Mass |
5000
kg |
250000
kg |
0,4
% |
0,4
% |
Master meter |
Internal procedure D1-2 D1-9 |
5000...250000 kg Watertemperature 4...85°C |
|
FLOW-Aarhus |
Flow |
Velocity transducers |
Velocity (air) (Bestemmelse af fringespacing for Laser Doppler Velocimetry (LDV) udstyr ved en hastighed på 8,9 m/s
)
|
8
m/s |
9
m/s |
0,1
% |
0,1
% |
Flywheel |
Internal procedure D1-19 |
Determination of fringe spacing of LDV equipment at velocity 8.9 m/s |
|
LUFT-Aarhus |
Flow |
Anemometers |
Velocity (air) |
0,05
m/s |
0,7
m/s |
0,02
m/s |
0,02
m/s |
Towing tank |
Internal procedure D1-2 |
Only thermo anemometers |
|
LUFT-Aarhus |
Flow |
Anemometers |
Velocity (air) |
0,3
m/s |
10
m/s |
0,0083
m/s |
0,046
m/s |
Wind tunnel and LDV |
Internal procedure D1-6 |
Max. blockage area: 200cm2 |
|
LUFT-Aarhus |
Flow |
Anemometers |
Velocity (air) |
10
m/s |
30
m/s |
0,046
m/s |
0,15
m/s |
Wind tunnel and LDV |
Internal procedure D1-6 |
Max. blockage area: 200cm2 |
|
LUFT-Aarhus |
Flow |
Flowmeters |
Volume flow rate l/s (0,4 - 2 l/min
)
|
6,7
ml/s |
33,3
ml/s |
0,33
ml/s |
1
ml/s |
Laminar flow element |
Internal procedure D2-1 |
With atmospheric air max pressure 7 bar |
|
LUFT-Aarhus |
Flow |
Flowmeters |
Volume flow rate l/s (1,5 - 33 l/min
)
|
0,025
l/s |
0,505
l/s |
2,08
ml/s |
17,3
ml/s |
Laminar flow element |
Internal procedure D2-1 |
With atmospheric air max pressure 7 bar |
|
LUFT-Aarhus |
Flow |
Flowmeters |
Volume flow rate l/s (10 - 300 l/min
)
|
0,167
l/s |
5
l/s |
0,011
l/s |
0,2
l/s |
Laminar flow element |
Internal procedure D2-1 |
With atmospheric air max pressure 7 bar |
|
LUFT-Aarhus |
Pressure and vacuum |
Manometers |
Pressure |
1
Pa |
50
Pa |
0,05
Pa |
0,05
Pa |
Dead weight tester, Piston/cylinder on electronic weighing machine |
Internal procedure P02.1, P03.1, P03.2 or P05.1 |
Micromanometers, differential pressure or gauge pressure. |
|
TRYK-Aarhus |
Pressure and vacuum |
Manometers |
Pressure |
50
Pa |
3100
Pa |
0,05
Pa |
0,3
Pa |
Dead weight tester, Piston/cylinder on electronic weighing machine |
Internal procedure P02.1, P03.1, P03.2 or P05.1 |
Micromanometers, differential pressure or gauge pressure. |
|
TRYK-Aarhus |
Pressure and vacuum |
Barometers |
Pressure |
75
kPa |
115
kPa |
4,7
Pa |
5,3
Pa |
Electronic barometer |
Internal procedure P06.1 |
Barometric pressure |
|
TRYK-Aarhus |
Pressure and vacuum |
Manometers |
Pressure |
2
kPa |
4,5
kPa |
0,35
Pa |
0,35
Pa |
dead weight tester |
Internal procedure P02.1, P03.1, P03.2 or P05.1 |
Gauge pressure, Pressure medium: gas |
|
TRYK-Aarhus |
Pressure and vacuum |
Manometers |
Pressure |
4,5
kPa |
20
kPa |
75
ppm |
75
ppm |
dead weight tester |
Internal procedure P02.1, P03.1, P03.2 or P05.1 |
Gauge pressure, Pressure medium: gas |
|
TRYK-Aarhus |
Pressure and vacuum |
Manometers |
Pressure |
20
kPa |
20
MPa |
36
ppm |
36
ppm |
dead weight tester |
Internal procedure P02.1, P03.1, P03.2 or P05.1 |
Gauge pressure, Pressure medium: gas |
|
TRYK-Aarhus |
Pressure and vacuum |
Manometers |
Pressure |
0,4
MPa |
120
MPa |
58
Pa |
4,6
kPa |
Dead weight tester |
Internal procedure P02.1, P03.1, P03.2 or P05.1 |
Gauge pressure, pressure medium: oil |
|
TRYK-Aarhus |
Pressure and vacuum |
Manometers |
Pressure |
0
kPa |
80
kPa |
39
Pa |
23
Pa |
Digital Pressure Controller |
Internal procedure P02.1, P03.1, P03.2 or P05.1 |
Absolute pressure, pressure medium: gas |
|
TRYK-Aarhus |
Pressure and vacuum |
Manometers |
Pressure |
80
kPa |
115
kPa |
4,7
Pa |
5,3
Pa |
Electronic barometer |
Internal procedure P02.1, P03.1, P03.2 or P05.1 |
Absolute pressure, pressure medium: gas |
|
TRYK-Aarhus |
Pressure and vacuum |
Manometers |
Pressure |
115
kPa |
120
kPa |
5,2
Pa |
5,3
Pa |
Dead weight tester + barometer |
Internal procedure P02.1, P03.1, P03.2 or P05.1 |
Absolute pressure, pressure medium: gas |
|
TRYK-Aarhus |
Pressure and vacuum |
Manometers |
Pressure |
120
kPa |
4,1
MPa |
15
Pa |
146
Pa |
Dead weight tester + barometer |
Internal procedure P02.1, P03.1, P03.2 or P05.1 |
Absolute pressure, pressure medium: gas |
|
TRYK-Aarhus |
Pressure and vacuum |
Manometers |
Pressure |
4,1
MPa |
20
MPa |
36
ppm |
36
ppm |
Dead weight tester + barometer |
Internal procedure P02.1, P03.1, P03.2 or P05.1 |
Absolute pressure, pressure medium: gas |
|
TRYK-Aarhus |
Pressure and vacuum |
Manometers |
Pressure |
500
kPa |
120
MPa |
64
Pa |
4,6
kPa |
Dead weight tester + barometer |
Internal procedure P02.1, P03.1, P03.2 or P05.1 |
Absolute pressure, Pressure medium: oil |
|
TRYK-Aarhus |
Pressure and vacuum |
Manometers |
Pressure (Differenstryk
)
|
0
Pa |
8
MPa |
10
Pa |
285
Pa |
Dead weight tester |
Internal procedure P04.1 or P05.1 |
Differential pressure with a line pressure up to 19,5 MPa. |
|
TRYK-Aarhus |
Pressure and vacuum |
Manometers |
Pressure (Differenstryk
)
|
8
MPa |
20
MPa |
285
Pa |
680
Pa |
Dead weight tester |
Internal procedure P04.1 or P05.1 |
Differential pressure with a line pressure up to 19,5 MPa. |
|
TRYK-Aarhus |
Pressure and vacuum |
Deadweight testers |
Pressure |
4,5
kPa |
20
kPa |
75
ppm |
75
ppm |
Dead weight tester |
Internal procedure D1-1.1 or D1-1.2 |
Determination of effective area or correction factor, pressure medium: gas |
|
TRYK-Aarhus |
Pressure and vacuum |
Deadweight testers |
Pressure |
20
kPa |
20
MPa |
41
ppm |
41
ppm |
Dead weight tester |
Internal procedure D1-1.1 or D1-1.2 |
Determination of effective area or correction factor, pressure medium: gas |
|
TRYK-Aarhus |
Pressure and vacuum |
Deadweight testers |
Pressure |
400
kPa |
1,2
MPa |
60
Pa |
84
Pa |
Dead weight tester |
Internal procedure D1-1.3 or D1-1.4 |
Determination of effective area or correction factor, pressure medium: oil |
|
TRYK-Aarhus |
Pressure and vacuum |
Deadweight testers |
Pressure |
1,2
MPa |
8
MPa |
60
ppm |
60
ppm |
Dead weight tester |
Internal procedure D1-1.3 or D1-1.4 |
Determination of effective area or correction factor, pressure medium: oil |
|
TRYK-Aarhus |
Pressure and vacuum |
Deadweight testers |
Pressure |
8
MPa |
120
MPa |
45
ppm |
45
ppm |
Dead weight tester |
Internal procedure D1-1.3 or D1-1.4 |
Determination of effective area or correction factor, pressure medium: oil |
|
TRYK-Aarhus |
Mass |
Weights |
Mass |
100
g |
2
kg |
0,000068
g |
0,00092
g |
Masses |
Internal procedure P08.1 |
Vejning ved substitution |
|
TRYK-Aarhus |
Mass |
Weights |
Mass |
2
kg |
20
kg |
0,00092
g |
0,01
g |
Masses |
Internal procedure P08.1 |
Vejning ved substitution |
|
TRYK-Aarhus |
Mass |
Weighing instruments |
Mass |
0,001
g |
1
g |
0,000027
g |
0,00013
g |
Weights |
Auto D1-10.1 |
Calibration of non-automatic Balances |
X |
AUTO-On site |
Mass |
Weighing instruments |
Mass |
0,001
kg |
1
kg |
0,00013
g |
0,007
g |
Weights |
Auto D1-10.1 |
Calibration of non-automatic Balances |
X |
AUTO-On site |
Mass |
Weighing instruments |
Mass |
1
kg |
10
kg |
0,007
g |
0,07
g |
Weights |
Auto D1-10.1 |
Calibration of non-automatic Balances |
X |
AUTO-On site |
Mass |
Weighing instruments |
Mass |
10
kg |
60
kg |
0,07
g |
3,4
g |
Weights |
Auto D1-10.1 |
Kalibrering af non-automatic Balances |
X |
AUTO-On site |
Temperature |
Thermometers |
Temperature |
-189,3442
°C |
0,01
°C |
0,001
°C |
0,001
°C |
Fixed points: Argon TP, Mercury TP, Water TP |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
-38,8344
°C |
29,7646
°C |
0,001
°C |
0,001
°C |
Fixed points: Mercury TP, Water TP, Gallium MP |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
0,01
°C |
29,7646
°C |
0,001
°C |
0,001
°C |
Fixed points: Water TP, Gallium MP |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
0,01
°C |
156,5985
°C |
0,001
°C |
0,0025
°C |
Fixed points: Water TP, Indium MP |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
0,01
°C |
231,928
°C |
0,001
°C |
0,0025
°C |
Fixed Points: Water TP, Indium TP and Tin (TP) |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
0,01
°C |
419,527
°C |
0,001
°C |
0,0025
°C |
Fixed Points: Water TP, Tin FP and Zinc FP |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
0,01
°C |
660,323
°C |
0,001
°C |
0,005
°C |
Fixed Points: Water TP, Tin FP, Zinc FP and Aluminium FP |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
-189,3442
°C |
-189,3442
°C |
0,00044
°C |
0,00044
°C |
Fixed Point: Argon TP |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
-38,8344
°C |
-38,8344
°C |
0,0005
°C |
0,0005
°C |
Fixed Point: Mercury TP |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
0,01
°C |
0,01
°C |
0,00011
°C |
0,00011
°C |
Fixed Point: Water TP |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
29,7646
°C |
29,7646
°C |
0,0006
°C |
0,0006
°C |
Fixed Point: Gallium MP |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
156,5985
°C |
156,5985
°C |
0,0021
°C |
0,0011
°C |
Fixed Point: Indium TP |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
231,928
°C |
231,928
°C |
0,002
°C |
0,002
°C |
Fixed Point: Tin FP |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
419,527
°C |
419,527
°C |
0,002
°C |
0,002
°C |
Fixed Point: Zinc FP |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
660,323
°C |
660,323
°C |
0,0048
°C |
0,0048
°C |
Fixed Point: Aluminium FP |
Internal procedure D1-1.2 |
Standard Platinum Resistance Thermometer (SPRT) |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
0
°C |
0
°C |
0,0025
°C |
0,0025
°C |
Ice Melting Point |
Internal procedure: D1-2.1, D1-2.2, D1-3.1, D1-4.1 |
Other Thermometers than SPRT's |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
-180
°C |
-80
°C |
0,048
°C |
0,024
°C |
Reference Thermometer |
Internal procedure D1-2.1, D1-2.2, D1-3.1, D1-4.1 |
Other thermometers than SPRT's by comparison in a cryostat |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
40
°C |
80
°C |
0,007
°C |
0,008
°C |
Reference Thermometer |
Internal procedure D1-2.1, D1-2.2, D1-3.1, D1-4.1 |
Other thermometers than SPRT's by comparison in liquid bath |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
-90
°C |
30
°C |
0,0045
°C |
0,0036
°C |
Reference Thermometer |
Internal procedure: D1-2.1, D1-2.2, D1-3.1, D1-4.1 |
Other thermometers than SPRT's by comparison in liquid bath |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
30
°C |
40
°C |
0,0063
°C |
0,0063
°C |
Reference Thermometer |
Internal procedure D1-2.1, D1-2.2, D1-3.1, D1-4.1 |
Other thermometers than SPRT's by comparison in liquid bath |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
40
°C |
200
°C |
0,0033
°C |
0,0056
°C |
Reference Thermometer |
Internal procedure: D1-2.1, D1-2.2, D1-3.1, D1-4.1 |
Other thermometers than SPRT's by comparison in liquid bath |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
200
°C |
550
°C |
0,0087
°C |
0,0089
°C |
Reference Thermometer |
Internal procedure: D1-2.1, D1-2.2, D1-3.1, D1-4.1 |
Other thermometers than SPRT's by comparison in liquid bath |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
200
°C |
340
°C |
0,45
°C |
0,45
°C |
Reference Thermometer |
Internal procedure: D1-2.1, D1-2.2, D1-3.1, D1-4.1 |
Other thermometers than SPRT's by comparison in a horisontal tube furnace |
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature |
340
°C |
1100
°C |
0,45
°C |
0,86
°C |
Reference Thermometer |
Internal procedure: D1-2.1, D1-2.2, D1-3.1, D1-4.1 |
Other thermometers than SPRT's by comparison in a horisontal tube furnace |
|
TERM-Aarhus |
Temperature |
Thermostat baths and ovens |
Temperature |
-100
°C |
-45
°C |
0,026
°C |
0,009
°C |
Reference Thermometer |
Internal procedure: D1-5.1 |
Dry Block Calibrators |
|
TERM-Aarhus |
Temperature |
Thermostat baths and ovens |
Temperature |
-45
°C |
155
°C |
0,009
°C |
0,009
°C |
Reference Thermometer |
Internal procedure: D1-5.1 |
Dry Block Calibrators and Micro-Baths |
|
TERM-Aarhus |
Temperature |
Thermostat baths and ovens |
Temperature |
155
°C |
350
°C |
0,047
°C |
0,047
°C |
Reference Thermometer |
Intern procedure: D1-5.1 |
Dry Block Calibrators and Micro-Baths |
|
TERM-Aarhus |
Temperature |
Thermostat baths and ovens |
Temperature |
350
°C |
660
°C |
0,13
°C |
0,13
°C |
Reference Thermometer |
Intern procedure: D1-5.1 |
Dry Block Calibrators and Micro-Baths |
|
TERM-Aarhus |
Temperature |
Thermostat baths and ovens |
Temperature |
660
°C |
1200
°C |
0,7
°C |
1,2
°C |
Reference Thermometer |
Internal procedure: D1-5.1 |
Dry Block Calibrators |
|
TERM-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Voltage (DC) |
0
V |
0,2
V |
0,35
μV |
0,75
μV |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Voltage (DC) |
0,2
V |
2
V |
1,8
μV |
6,1
μV |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Voltage (DC) |
2
V |
20
V |
10
μV |
68
μV |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Voltage (DC) |
20
V |
200
V |
0,24
mV |
0,89
mV |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Voltage (AC) (40 Hz - 10 kHz
)
|
200
V |
1000
V |
58
ppm |
230
ppm |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Voltage (DC) |
200
V |
1000
V |
2,8
mV |
5,6
mV |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Voltage (AC) (10 Hz - 10 kHz
)
|
0,1
V |
0,2
V |
61
ppm |
200
ppm |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Voltage (AC) (10 kHz - 100 kHz
)
|
0,1
V |
0,2
V |
180
ppm |
410
ppm |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Voltage (AC) (10 Hz - 10 kHz
)
|
0,2
V |
2
V |
39
ppm |
180
ppm |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Voltage (AC) (10 kHz - 100 kHz
)
|
0,2
V |
2
V |
120
ppm |
0,65
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Voltage (AC) (10 Hz - 10 kHz
)
|
2
V |
20
V |
40
ppm |
200
ppm |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Voltage (AC) (10 kHz - 100 kHz
)
|
2
V |
20
V |
170
ppm |
0,66
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Voltage (AC) (10 Hz - 10 kHz
)
|
20
V |
200
V |
43
ppm |
240
ppm |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Voltage (AC) (10 kHz - 100 kHz
)
|
20
V |
200
V |
120
ppm |
0,65
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
0
Ω |
2
Ω |
6,8
μΩ |
26
μΩ |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
2
Ω |
20
Ω |
42
μΩ |
0,16
mΩ |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
20
Ω |
200
Ω |
0,32
mΩ |
1,5
mΩ |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
200
Ω |
2
kΩ |
3,2
mΩ |
15
mΩ |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
2
kΩ |
20
kΩ |
32
mΩ |
0,15
Ω |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
20
kΩ |
200
kΩ |
0,32
Ω |
1,5
Ω |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
200
kΩ |
2
MΩ |
4
Ω |
16
Ω |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
2
MΩ |
20
MΩ |
18
Ω |
0,34
kΩ |
Digital Multimeter |
Internal procedure D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
20
MΩ |
200
MΩ |
16
kΩ |
23
kΩ |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
0
μA |
200
μA |
0,68
nA |
2,1
nA |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
200
μA |
2
mA |
7,9
nA |
21
nA |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
2
mA |
20
mA |
82
nA |
0,24
μA |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
20
mA |
200
mA |
1,9
μA |
8,7
μA |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
200
mA |
2
A |
35
μA |
0,21
mA |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (10 Hz - 3 kHz
)
|
200
mA |
2
A |
700
ppm |
0,18
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (3 kHz - 30 kHz
)
|
200
mA |
2
A |
700
ppm |
0,31
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (10 Hz - 3 kHz
)
|
2
A |
20
A |
900
ppm |
0,35
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (3 Hz - 10 kHz
)
|
2
A |
20
A |
0,26
% |
0,35
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (10 Hz - 3 kHz
)
|
100
μA |
200
μA |
580
ppm |
680
ppm |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (3 kHz - 30 kHz
)
|
100
μA |
200
μA |
580
ppm |
0,42
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
2
A |
20
A |
0,81
mA |
4,7
mA |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (10 Hz - 3 kHz
)
|
200
μA |
2
mA |
380
ppm |
0,13
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (3 kHz - 30 kHz
)
|
200
μA |
2
mA |
380
ppm |
0,51
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (3 kHz - 30 kHz
)
|
2
mA |
20
mA |
380
ppm |
0,42
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (10 Hz - 3 kHz
)
|
2
mA |
20
mA |
380
ppm |
0,13
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (10 Hz - 3 kHz
)
|
20
mA |
200
mA |
580
ppm |
0,15
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) (3 kHz - 30 kHz
)
|
20
mA |
200
mA |
580
ppm |
0,42
% |
Digital Multimeter |
Internal procedure: D1-7.1, D1-7.2, D1-7.3 |
|
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Power (45 Hz - 65 Hz
)
|
82,5
mW |
100
mW |
0,59
% |
0,12
% |
Electrical Calibrator |
Internal procedure: 4330 |
PF = 1 |
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Power (45 Hz - 65 Hz
)
|
0,1
W |
3000
W |
0,12
% |
0,14
% |
Electrical Calibrator |
Internal procedure: 4330 |
PF = 1 |
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Power (45 Hz - 65 Hz
)
|
3
kW |
21
kW |
0,14
% |
0,21
% |
Electrical Calibrator |
Internal procedure: 4330 |
PF = 1 |
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Power (45 Hz - 65 Hz
)
|
82,5
mW |
100
mW |
0,58
% |
0,16
% |
Electrical Calibrator |
Internal procedure: 4330 |
PF < 1; U(CMC) relative to VA |
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Power (45 Hz - 65 Hz
)
|
0,1
W |
3000
W |
0,15
% |
0,15
% |
Electrical Calibrator |
Internal procedure: 4330 |
PF < 1; U(CMC) relative to VA |
|
El-Aarhus |
Electricity DC and LF |
Electrical equipment (AC) |
Power (45 Hz - 65 Hz
)
|
3
kW |
21
kW |
0,24
% |
0,24
% |
Electrical Calibrator |
Internal procedure: 4330 |
PF < 1; U(CMC) relative to VA |
|
El-Aarhus |
Time and frequency |
Time and frequence meas. eq. |
Timeinterval |
1
μs |
1
s |
6,6
ns |
7,1
μs |
counter |
Internal procedure: D1-9.3 |
|
|
El-Aarhus |
Time and frequency |
Time and frequence meas. eq. |
Timeinterval |
1
s |
1
ks |
7,1
μs |
7,1
ms |
counter |
Internal procedure: D1-9.3 |
|
|
El-Aarhus |
Time and frequency |
Time and frequence meas. eq. |
Frequency |
100
mHz |
10
MHz |
12
pHz |
1,2
mHz |
GPSDO and counter |
Internal procedure: D1-9.1, D1-9.2 |
|
|
El-Aarhus |
Time and frequency |
Time and frequence meas. eq. |
Frequency |
10
MHz |
225
MHz |
1,2
mHz |
26
mHz |
GPSDO and counter |
Internal procedure: D1-9.1, D1-9.2 |
|
|
El-Aarhus |
Time and frequency |
Time and frequence meas. eq. |
Period |
4,4
ns |
1
μs |
0,00052
fs |
0,12
fs |
GPSDO and counter |
Internal procedure: D1-9.1, D1-9.2 |
|
|
El-Aarhus |
Time and frequency |
Time and frequence meas. eq. |
Period |
1
μs |
10
s |
0,12
fs |
1,2
ns |
GPSDO and counter |
Internal procedure: D1-9.1, D1-9.2 |
|
|
El-Aarhus |
Temperature |
Pyrometers and IR equipments |
Temperature (suface) |
-40
°C |
80
°C |
0,3
°C |
0,3
°C |
Reference thermometer, blackbody |
Intern procedure D1-15 |
|
|
TERM-Aarhus |
Time and frequency |
Time and frequence meas. eq. |
Timeinterval |
30
s |
172800
s |
0,038
s |
0,043
s |
GPS / WaveForm Generator / Counter |
Internal procedure Calibration of stopwatches (ID:9777) |
Time Interval (Calibration of Stopwatches). |
|
El-Aarhus |
Temperature |
Pyrometers and IR equipments |
Temperature (suface) |
80
°C |
270
°C |
0,6
°C |
0,6
°C |
Reference thermometer, blackbody |
Intern procedure D1-15 |
|
|
TERM-Aarhus |
Temperature |
Thermometers |
Temperature (air) |
-40
°C |
0
°C |
0,2
°C |
0,13
°C |
Digital thermometer |
Internal procedure D1-6.1 |
Air temperature |
|
FUGT-Aarhus |
Temperature |
Thermometers |
Temperature (air) |
0
°C |
70
°C |
0,028
°C |
0,049
°C |
Digital thermometer |
Internal procedure D1-6.1 |
Air temperature |
|
FUGT-Aarhus |
Temperature |
Thermometers |
Temperature (air) |
70
°C |
90
°C |
0,17
°C |
0,25
°C |
Digital thermometer |
Internal procedure D1-6.1 |
Air temperature |
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Relative humidity of air (5 - 20 °C
)
|
5
%rh |
10
%rh |
0,16
%rh |
0,25
%rh |
Dew point instrument |
Internal procedure D1-14 |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Relative humidity of air (20 - 30 °C
)
|
0,2
%rh |
10
%rh |
0,12
%rh |
0,2
%rh |
Dew point instrument |
Internal procedure D1-14 |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Relative humidity of air (30 - 50 °C
)
|
5
%rh |
10
%rh |
0,14
%rh |
0,17
%rh |
Dew point instrument |
Internal procedure D1-14 |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Relative humidity of air (0 - 70 °C
)
|
10
%rh |
95
%rh |
0,14
%rh |
0,48
%rh |
Dew point instrument |
Internal procedure D1-6.1 |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Relative humidity of air (70 - 80 °C
)
|
10
%rh |
66
%rh |
0,34
%rh |
1,1
%rh |
Dew point instrument |
Internal procedure |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Relative humidity of air (80 - 90 °C
)
|
10
%rh |
45
%rh |
0,34
%rh |
0,73
%rh |
Dew point instrument |
Internal procedure D1-6.1 |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Dewpoint temperature |
-75
°C |
-50
°C |
0,19
°C |
0,056
°C |
Dew point instrument |
Internal procedure D1-14 |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Dewpoint temperature |
-50
°C |
70
°C |
0,056
°C |
0,085
°C |
Dew point instrument |
Internal procedure D1-6.1 and D1-14 |
|
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Dewpoint temperature |
-75
°C |
-50
°C |
0,2
°C |
0,14
°C |
Dewpoint instrument |
Internal procedure D1-14 |
Pressure dewpoint: 0 - 7 bar |
|
FUGT-Aarhus |
Humidity |
Hygrometers |
Dewpoint temperature |
-50
°C |
10
°C |
0,14
°C |
0,14
°C |
Dewpoint instrument |
Internal procedure D1-14 |
Pressure dewpoint: 0 - 7 bar |
|
FUGT-Aarhus |
Humidity |
Moisture meter |
Water content, g(H2O)/g(dry) |
8
% |
8
% |
0,5
% |
0,5
% |
Balance |
Internal procedure: 8220 |
Wood type: spruce, fir not impregnated |
|
FUGT-Aarhus |
Humidity |
Moisture meter |
Water content, g(H2O)/g(dry) |
17
% |
17
% |
0,8
% |
0,8
% |
Balance |
Internal procedure: 8220 |
Wood type: spruce, fir not impregnated |
|
FUGT-Aarhus |
Humidity |
Moisture meter |
Water content, g(H2O)/g(dry) |
20
% |
20
% |
0,9
% |
0,9
% |
Balance |
Internal procedure: 8220 |
Wood type: spruce, fir not impregnated |
|
FUGT-Aarhus |
Humidity |
Moisture meter |
Water content, g(H2O)/g(dry) |
26,7
% |
26,7
% |
1,1
% |
1,1
% |
Balance |
Internal procedure: 8220 |
Wood type: spruce, fir not impregnated |
|
FUGT-Aarhus |
Chemical |
Gas concentrations |
Volume % (CO2
)
|
42
ppm |
48
vol.% |
2
% |
2,8
% |
Reference gas and flow meter |
Internal Procedure D1-12 |
|
|
FUGT-Aarhus |
Temperature |
Temperature Gauges |
Temperature |
-45
°C |
155
°C |
0,14
°C |
0,14
°C |
Dry block calibrator |
Internal Procedure Auto D1-2.1 |
Digitaltermometer, opløsning 0,01°C |
X |
AUTO-On site |
Pressure and vacuum |
Manometers |
Pressure |
-100
kPa |
0
kPa |
1,5
kPa |
1,5
kPa |
Pressure Calibrator |
Internal procedure Auto D1-3.1, DS/EN 837 |
Vakuum. Opløsning 1 kPa |
X |
AUTO-On site |
Pressure and vacuum |
Manometers |
Pressure |
0
MPa |
2
MPa |
1,7
kPa |
1,7
kPa |
Pressure Calibrator |
Internal procedure
Auto D1-3.1 |
Overtryk. Opløsning 1 kPa |
X |
AUTO-On site |
Pressure and vacuum |
Manometers |
Pressure |
0
Pa |
7
MPa |
2
kPa |
9
kPa |
Pressure Calibrator |
Internal procedure Auto D1-3.1 |
Overtryk. Opløsning 1 kPa |
X |
AUTO-On site |
Pressure and vacuum |
Manometers |
Pressure |
0
Pa |
30
MPa |
13
kPa |
40
kPa |
Pressure Calibrator |
Internal procedure Auto D1-3.1, DS/EN 837 |
Resolution 10 kPa |
X |
AUTO-On site |
Pressure and vacuum |
Manometers |
Pressure |
0
Pa |
70
MPa |
48
kPa |
48
kPa |
Pressure Calibrator |
Internal procedure
Auto D1-3.1 |
Overtryk. Opløsning 10 kPa |
X |
AUTO-On site |
Force and torque |
Torque measuring devices |
Torque |
1
Nm |
1100
Nm |
1,8
% |
0,62
% |
Torque Wrench Analyzer |
Internal procedure Auto D1-6.1, DS/EN ISO 6789-2:2017 |
|
X |
AUTO-On site |
Dimensional |
Caliper |
Length |
0
mm |
500
mm |
20
μm |
48
μm |
Gauge blocks |
Internal procedure
Auto D1-8.1 |
Opløsning 0,01 mm |
X |
AUTO-On site |
Dimensional |
Micrometer (all types) |
Length |
0
mm |
25
mm |
3,1
μm |
14,2
μm |
Gauge blocks |
Internal procedure Auto D1-8.1 |
Opløsning 0,01 mm |
X |
AUTO-On site |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
25
mm |
2,8
μm |
3,3
μm |
Dial Gage Tester |
Internal procedure Auto D1-8.1 |
Opløsning: 0,001 mm |
X |
AUTO-On site |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
300
mm |
3,1
μm |
9,5
μm |
Lenght Transducer (Digital Dial Gauge) and Gauge Blocks |
Intern procedure Auto D1-8.1 |
Including Build-In Lenght Tranducers. Resolution 0,001 mm.
|
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (DC) |
Voltage (DC) |
0
V |
1100
V |
5,7
μV |
154
mV |
Multifunctioncalibrator |
Internal procedure
Auto D1.5.1 |
Opløsning: 0,001mV - 0,01V |
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
0
A |
20
A |
0,02
nA |
10
mA |
Multifunctioncalibrator |
Internal procedure
Auto D1.5.2 |
Opløsning: 0,01µA - 1mA |
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
20
A |
1000
A |
0,05
A |
2,5
A |
Multifunctioncalibrator |
Internal procedure Auto D1.5.2 |
|
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
0
Ω |
400
MΩ |
23
mΩ |
0,34
MΩ |
Multifunctioncalibrator |
Internal procedure
Auto D1.5.3 |
Opløsning: 0,001 ohm - 0,01Mohm |
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (AC) |
Voltage (AC) |
0
V |
800
V |
0,45
mV |
742
mV |
Multifunctioncalibrator |
Internal procedure
Auto D1.5.4 |
Opløsning: 0,001mV - 0,01V |
X |
AUTO-On site |
Electricity DC and LF |
Tachometers |
RPM |
600
rpm |
6000
rpm |
2,3
rpm |
2,9
rpm |
Enginesimulator |
Internal procedure
Auto D1.5.5 |
Generering. Opløsning: 1 rpm |
X |
AUTO-On site |
Electricity DC and LF |
Tachometers |
RPM |
0
rpm |
90000
rpm |
2,1
rpm |
56,1
rpm |
Tachometer |
Internal procedure
Auto D1.5.5 |
Måling, opløsning: 1rpm |
X |
AUTO-On site |
Electricity DC and LF |
Motor testers |
Cam angle |
0
° |
100
° |
0,12
° |
0,12
° |
Enginesimulator |
Internal procedure
Auto D1.5.6 |
Opløsning: 0,1% |
X |
AUTO-On site |
Electricity DC and LF |
Motor testers |
|
0
° |
50
° |
0,29
° |
0,29
° |
Enginesimulator |
Intern procedure
Auto D1.5.6 |
opløsning: 0,1° |
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) |
0
A |
20
A |
0,44
μA |
51
mA |
Multifunctioncalibrator |
Internal procedure Auto D1.5.7 |
|
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) |
20
A |
200
A |
125
mA |
700
mA |
Multifunctioncalibrator |
Internal procedure Auto D1.5.7 |
|
X |
AUTO-On site |
Force and torque |
Brake testers |
Force |
1
kN |
31
kN |
32,1
N |
302
N |
weigths and weigtharm |
Internal procedure
Auto D1.7.1 |
Opløsning: 2daN |
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) |
200
A |
1000
A |
1
A |
3,5
A |
Multifunctioncalibrator |
Internal procedure Auto D1.5.7 |
|
X |
AUTO-On site |
Force and torque |
Torque wrenches |
Torque |
1
Nm |
1100
Nm |
1,8
% |
0,62
% |
Torque Wrench Analyzer |
Internal procedure Auto D1-6.1, DS/EN ISO 6789:2003 |
|
X |
AUTO-On site |
Force and torque |
Torque transducers |
Torque |
0,2
Nm |
10
Nm |
1
% |
0,5
% |
Transfer Torque Wrench |
Internal Procedure Auto D1.9.1 Level 1 |
|
|
AUTO-Aarhus |
Force and torque |
Torque transducers |
Torque |
10
Nm |
100
Nm |
0,5
% |
0,4
% |
Transfer torque wrench - TTS 100 Nm |
Internal procedure Auto D1.9.1 Level 1 |
|
|
AUTO-Aarhus |
Force and torque |
Torque transducers |
Torque |
100
Nm |
1100
Nm |
0,4
% |
0,4
% |
Transfer torque wrench - TTS 1000 Nm |
Internal procedure Auto D1.9.1 Level 1 |
|
|
AUTO-Aarhus |
Force and torque |
Torque transducers |
Torque |
10
Nm |
100
Nm |
0,8
% |
0,8
% |
Torque Transducer |
Internal procedure Auto D1.9.1 Level 2 |
|
X |
AUTO-Århus |
Force and torque |
Torque transducers |
Torque |
1
Nm |
10
Nm |
0,9
% |
0,8
% |
Torque Transducer |
Internal procedure Auto D1.9.1 - Level 2 |
|
X |
AUTO-Aarhus |
Force and torque |
Torque transducers |
Torque |
100
Nm |
1100
Nm |
0,8
% |
0,7
% |
Torque Transducer |
Internal procedure Auto D1.9.1 Level 2 |
|
X |
AUTO-Århus |
Time and frequency |
Time and frequence meas. eq. |
Frequency |
0
Hz |
10
MHz |
0,0006
Hz |
300
Hz |
Multifunctioncalibrator |
Internal procedure Auto D1.5.5 |
|
X |
AUTO-On site |
Time and frequency |
Time and frequence meas. eq. |
Frequency |
40
Hz |
400
Hz |
0,058
Hz |
0,58
Hz |
Installation testers and Insulation testers calibrator. |
Intern procedure D1-11.1 |
Resolution: : 0,1 Hz- 1Hz |
X |
AUTO-On site |
Force and torque |
Force transducers |
Force (0,001 - 1,3 kN
)
|
1
N |
1300
N |
0,05
% |
0,05
% |
Deadweight |
DS/EN ISO 376:2011 |
Calibration of Force Transducer Tension. Internal method K03. |
|
KRAFT-Aarhus |
Force and torque |
Force transducers |
Force |
0,1
kN |
1
kN |
0,09
% |
0,05
% |
Force Transducer |
DS/EN ISO 376:2011 |
Calibration of Force Transducer Tension. Internal method K03. |
|
KRAFT-Aarhus |
Force and torque |
Force transducers |
Force |
1
kN |
600
kN |
0,05
% |
0,05
% |
Force Transducer |
DS/EN ISO 376:2011 |
Calibration of Force Transducer Tension. Internal method K03. |
|
KRAFT-Aarhus |
Force and torque |
Force transducers |
Force (0,001 - 1,3 kN
)
|
1
N |
1300
N |
0,05
% |
0,05
% |
Deadweight |
DS/EN ISO 376:2011 |
Calibration of Force Transducers Compression. Internal method K03. |
|
KRAFT-Aarhus |
Force and torque |
Force transducers |
Force |
0,1
kN |
100
kN |
0,09
% |
0,05
% |
Force Transducers |
DS/EN ISO 376:2011 |
Calibration of Force Transducer Compression. Internal method K03. |
|
KRAFT-Aarhus |
Force and torque |
Force transducers |
Force |
100
kN |
1000
kN |
0,05
% |
0,08
% |
Force Transducers |
DS/EN ISO 376:2011 |
Calibration of Force Transducer Compression. Internal method K03. |
|
KRAFT-Aarhus |
Force and torque |
Force transducers |
Force |
1000
kN |
2000
kN |
0,08
% |
0,17
% |
Force Transducers |
DS/EN ISO 376:2011 |
Calibration of Force Transducer Compression. Internal method K03. |
|
KRAFT-Aarhus |
Force and torque |
Force transducers |
Force |
2000
kN |
3000
kN |
0,17
% |
0,18
% |
Force Transducers |
DS/EN ISO 376:2011 |
Calibration of Force Transducer Compression. Internal method K03. |
|
KRAFT-Aarhus |
Force and torque |
Force transducers |
Force |
3000
kN |
5000
kN |
0,18
% |
0,18
% |
Force Transducers |
DS/EN ISO 376:2011 |
Calibration of Force Transducer Compression. Internal method K03. |
|
KRAFT-Aarhus |
Force and torque |
Material testing machines |
Force |
1
N |
600
kN |
0,4
% |
0,3
% |
Deadweights and Force Transducers |
EN ISO 7500-1:2018 |
Calibration of Force Equipment Tension based, including Hydraulic Jack's, Dynamometers ect. Internal method K04. |
X |
KRAFT-Aarhus |
Force and torque |
Material testing machines |
Force |
1
N |
25
N |
0,4
% |
0,4
% |
Deadweights and Force Transducers |
EN ISO 7500-1:2018; EN 12390-4:2019 |
Calibration of compression testing machines, including hydraulic jack's, dynamometers ect. Internal method K01 & K04. |
X |
KRAFT-Aarhus |
Force and torque |
Material testing machines |
Force |
25
N |
1
kN |
0,4
% |
0,3
% |
Deadweights and Force Transducers |
EN ISO 7500-1:2018; EN 12390-4:2019 |
Calibration of compression testing machines, including hydraulic jack's, dynamometers ect. Internal method K01 & K04. |
X |
KRAFT-Aarhus |
Force and torque |
Material testing machines |
Force |
1
kN |
5000
kN |
0,3
% |
0,3
% |
Deadweights and Force Transducers |
EN ISO 7500-1:2018; EN 12390-4:2019 |
Calibration of compression testing machines, including hydraulic jack's, dynamometers ect. Internal method K01 & K04. |
X |
KRAFT-Aarhus |
Force and torque |
Material testing machines |
Force |
1
N |
1300
N |
0,25
% |
0,25
% |
Force Transducers |
EN ISO 7500-1:2018 |
Calibration of Force Equipment Tension & Compression, including Dynamometers ect. Internal metod K04. |
X |
KRAFT-Aarhus |
Hardness |
Hardness testers |
Hardness (Brinell) |
20
HB |
5000
HB |
11
% |
1
% |
Force Transducers, Geometrical Scale, Stop Watch. |
DS/EN ISO 6506-2:2018 |
Calibration of Hardness Testers, Direct Method. Except calibration of the endenter. For endentions > 0,2 mm. Internal method K05.1. |
X |
KRAFT-Aarhus |
Hardness |
Hardness testers |
Hardness (Vickers) |
50
HV |
900
HV |
2,5
% |
2,5
% |
Force Transducers, Geometrical Scale, Stop Watch. |
DS/EN ISO 6507-2:2018 |
Calibration of Hardness Testers. Direct Method. Except calibration of the endenter. For endentions > 0,2 mm. Internal method K05.2.
|
X |
KRAFT-Aarhus |
Hardness |
Hardness testers |
dBHardness (Rockwell) |
20
HR |
100
HR |
1,2
% |
0,4
% |
Force Transducers, Gauge blocks, Stop Watch. |
DS/EN ISO 6508-2:2015 |
Calibration of Hardness Testers. Direct Method. Except calibration of the endenter. Internal method K05.3. |
X |
KRAFT-Aarhus |
Force and torque |
Impact Testing Machines |
Impact |
10
J |
120
J |
1,5
J |
6,2
J |
Reference Impact material, Geometrical equipment, Force Transducer |
DS/EN ISO 148-2:2016 |
Calibration of Impact Testing Machines. Internal method K06. |
X |
KRAFT-Aarhus |
Force and torque |
Impact Testing Machines |
Impact |
120
J |
450
J |
6,2
J |
25
J |
Reference Impact material, Geometrical equipment, Force Transducer |
DS/EN ISO 148-2:2016 |
Calibration of Impact Testing Machines. Internal method K06. |
X |
KRAFT-Aarhus |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
0
Ω |
4
kΩ |
7,1
μΩ |
369,1
mΩ |
Milliohmmeter |
D1-12.1 |
|
X |
AUTO On-site |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
12
mm |
1,1
μm |
1,1
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Måleur, analog med opløsning 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
50
mm |
1,1
μm |
1,2
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Måleur, analog med opløsning 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
100
mm |
1,1
μm |
1,3
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Måleur, analog med opløsning 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
100
mm |
1,1
μm |
1,3
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Måleur, analog med opløsning 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
12
mm |
0,4
μm |
0,4
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Måleur, analog med opløsning 0,001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
50
mm |
0,4
μm |
0,6
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Måleur, analog med opløsning 0,001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
100
mm |
0,4
μm |
0,9
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Måleur, analog med opløsning 0,001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
100
mm |
5,8
μm |
5,8
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Måleur, digital med opløsning 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
12
mm |
0,7
μm |
0,7
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Måleur, digital med opløsning 0,001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
50
mm |
0,7
μm |
0,9
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Måleur, digital med opløsning 0,001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
100
mm |
0,7
μm |
1
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Måleur, digital med opløsning 0,001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Electronic probes |
Length |
0
mm |
10
mm |
0,3
μm |
0,3
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Elektroniske længdetransducer og El-tast, Opløsning 0,0001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Electronic probes |
Length |
0
mm |
50
mm |
0,3
μm |
0,6
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Elektroniske længdetransducer og El-tast, Opløsning 0,0001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Electronic probes |
Length |
0
mm |
100
mm |
0,3
μm |
0,9
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Elektroniske længdetransducer og El-tast, Opløsning 0,0001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Electronic probes |
Length |
0
mm |
10
mm |
0,3
μm |
0,3
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Elektroniske længdetransducer og El-tast. Opløsning 0,00001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Electronic probes |
Length |
0
mm |
50
mm |
0,3
μm |
0,6
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Elektroniske længdetransducer og El-tast. Opløsning 0,00001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Electronic probes |
Length |
0
mm |
100
mm |
0,3
μm |
0,9
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Elektroniske længdetransducer og El-tast. Opløsning 0,00001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
2
mm |
1,5
μm |
1,5
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Centrumindikatorer. Opløsning 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Dial gauges/centrumindicators |
Length |
0
mm |
1
mm |
0,4
μm |
0,4
μm |
Laserinterferometer og krafttransducer |
120.44.10.04 |
Centrumindikatorer. Opløsning 0,001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Caliper |
Length |
0
mm |
150
mm |
16
μm |
16
μm |
Måleklodser og referenceringe |
120.44.10.01 |
Skydelærer: elektronisk og mekanisk med ur. Opløsning 0,01 - 0,02 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Caliper |
Length |
0
mm |
300
mm |
16
μm |
18
μm |
Måleklodser og referenceringe |
120.44.10.01 |
Skydelærer: opløsning 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Caliper |
Length |
0
mm |
600
mm |
16
μm |
20
μm |
Måleklodser og referenceringe |
120.44.10.01 |
Skydelærer: opløsning 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Caliper |
Length |
0
mm |
1000
mm |
16
μm |
22
μm |
Måleklodser og referenceringe |
120.44.10.01 |
Skydelærer: elektronisk og mekanisk med ur. Opløsning 0,01 - 0,02 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Caliper |
Length |
0
mm |
150
mm |
18
μm |
18
μm |
Måleklodser og referenceringe |
120.44.10.01 |
Skydelærere: opløsning 0,05 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Caliper |
Length |
0
mm |
300
mm |
18
μm |
20
μm |
Måleklodser og referenceringe |
120.44.10.01 |
Skydelærere: opløsning 0,05 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Caliper |
Length |
0
mm |
600
mm |
18
μm |
22
μm |
Måleklodser og referenceringe |
120.44.10.01 |
Skydelærere: opløsning 0,05 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Micrometer (all types) |
Length |
0
mm |
25
mm |
3
μm |
3
μm |
Måleklodser, planparallelglas og vejecelle |
120.44.10.02 |
Mikrometerskrue: opløsning: 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Micrometer (all types) |
Length |
25
mm |
50
mm |
3
μm |
3,1
μm |
Måleklodser, planparallelglas og vejecelle |
120.44.10.02 |
Mikrometerskrue: opløsning: 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Micrometer (all types) |
Length |
50
mm |
75
mm |
3,1
μm |
3,3
μm |
Måleklodser, planparallelglas og vejecelle |
120.44.10.02 |
Mikrometerskrue: opløsning: 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Micrometer (all types) |
Length |
75
mm |
100
mm |
3,3
μm |
3,4
μm |
Måleklodser, planparallelglas og vejecelle |
120.44.10.02 |
Mikrometerskrue: opløsning: 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Micrometer (all types) |
Length |
100
mm |
200
mm |
3,4
μm |
4
μm |
Måleklodser, planparallelglas og vejecelle |
120.44.10.02 |
Mikrometerskrue: opløsning: 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Micrometer (all types) |
Length |
0
mm |
25
mm |
0,8
μm |
0,8
μm |
Måleklodser, planparallelglas og vejecelle |
120.44.10.02 |
Mikrometerskrue: opløsning: 0,001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Micrometer (all types) |
Length |
25
mm |
50
mm |
0,8
μm |
0,9
μm |
Måleklodser, planparallelglas og vejecelle |
120.44.10.02 |
Mikrometerskrue: opløsning: 0,001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Micrometer (all types) |
Length |
50
mm |
75
mm |
0,9
μm |
1,1
μm |
Måleklodser, planparallelglas og vejecelle |
120.44.10.02 |
Mikrometerskrue: opløsning: 0,001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Micrometer (all types) |
Length |
75
mm |
100
mm |
1,1
μm |
1,2
μm |
Måleklodser, planparallelglas og vejecelle |
120.44.10.02 |
Mikrometerskrue: opløsning: 0,001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Micrometer (all types) |
Length |
100
mm |
200
mm |
1,2
μm |
1,8
μm |
Måleklodser, planparallelglas og vejecelle |
120.44.10.02 |
Mikrometerskrue: opløsning: 0,001 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Internal micrometers |
Length |
25
mm |
75
mm |
1
μm |
1
μm |
Måleklodser og længdemålemaskine |
120.44.10.03 |
Grundenhed for mikrometerspærmål, opløsning 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Internal micrometers |
Length |
100
mm |
125
mm |
1,1
μm |
1,1
μm |
Måleklodser og længdemålemaskine |
120.44.10.03 |
Grundenhed for mikrometerspærmål, opløsning 0,01 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Internal micrometers |
Length |
10
mm |
25
mm |
0,3
μm |
0,3
μm |
Længdemålemaskine |
120.44.10.03 |
Forlængere for mikrometer spærmål |
|
M&K(GEOM)-Taastrup |
Dimensional |
Internal micrometers |
Length |
25
mm |
100
mm |
0,3
μm |
0,4
μm |
Længdemålemaskine |
120.44.10.03 |
Forlængere for mikrometer spærmål |
|
M&K(GEOM)-Taastrup |
Dimensional |
Internal micrometers |
Length |
100
mm |
300
mm |
0,4
μm |
0,9
μm |
Måleklodser og længdemålemaskine |
120.44.10.03 |
Forlængere for mikrometer spærmål |
|
M&K(GEOM)-Taastrup |
Dimensional |
Internal micrometers |
Length |
300
mm |
600
mm |
0,9
μm |
1
μm |
Måleklodser og længdemålemaskine |
120.44.10.03 |
Forlængere for mikrometer spærmål |
|
M&K(GEOM)-Taastrup |
Dimensional |
Setting standards |
Length |
0
mm |
50
mm |
0,3
μm |
0,3
μm |
Længdemålemaskine |
120.44.08.10 |
Indstillingsnormal |
|
M&K(GEOM)-Taastrup |
Dimensional |
Setting standards |
Length |
0
mm |
100
mm |
0,3
μm |
0,4
μm |
Længdemålemaskine |
120.44.08.10 |
Indstillingsnormal |
|
M&K(GEOM)-Taastrup |
Dimensional |
Setting standards |
Length |
0
mm |
300
mm |
0,3
μm |
0,7
μm |
Længdemålemaskine og måleklodser |
120.44.08.10 |
Indstillingsnormal |
|
M&K(GEOM)-Taastrup |
Dimensional |
Setting standards |
Length |
0
mm |
600
mm |
0,3
μm |
1
μm |
Længdemålemaskine og måleklodser |
120.44.08.10 |
Indstillingsnormal |
|
M&K(GEOM)-Taastrup |
Dimensional |
Measuring tapes (roller taper) |
Length |
0
mm |
50000
mm |
12
μm |
412
μm |
Laserinterferometer |
120.44.10.09 |
Målebånd |
|
M&K(GEOM)-Taastrup |
Dimensional |
Linear scales |
Length |
0
mm |
500
mm |
2,1
μm |
6,1
μm |
Målemikroskop |
120.44.10.10 |
Målestokke, linealer |
|
M&K(GEOM)-Taastrup |
Dimensional |
Linear scales |
Length |
1
mm |
5000
mm |
12
μm |
52
μm |
Laserinterferometer |
120.44.10.08 |
Målestokke, linealer |
|
M&K(GEOM)-Taastrup |
Dimensional |
Circometers |
Length |
100
mm |
4000
mm |
11,5
μm |
31
μm |
Laserinterferometer |
120.44.10.08 |
Cirkometer - diametermåling af runde emner |
|
M&K(GEOM)-Taastrup |
Dimensional |
Surface plates |
Length |
50
mm |
1000
mm |
1,7
μm |
4,8
μm |
3 Axes Measuring Machine |
120.44.06.07 |
Planes of stone or steel |
|
M&K(GEOM)-Taastrup/Aarhus |
Dimensional |
Height gauges |
Length |
0
mm |
1000
mm |
11,6
μm |
19,6
μm |
Måleklodser |
120.44.10.15 |
Opløsning: 10 µm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Height gauges |
Length |
0
mm |
1000
mm |
2
μm |
10
μm |
Måleklodser |
120.44.10.15 |
Opløsning: 1 µm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Setting standards |
Length |
0,01
mm |
10
mm |
0,5
μm |
0,5
μm |
Længdemålemaskine |
120.44.10.14 |
Bladsøger, kalibreringsfolie |
|
M&K(GEOM)-Taastrup |
Dimensional |
Ring gauges |
Length |
0
mm |
10
mm |
0,5
μm |
0,5
μm |
Længdemålemaskine og formtester (rundhed) |
120.44.08.02 |
Kontroldorne og tolerancedorne |
|
M&K(GEOM)-Taastrup |
Dimensional |
Ring gauges |
Length |
10
mm |
50
mm |
0,5
μm |
0,6
μm |
Længdemålemaskine og formtester (rundhed) |
120.44.08.02 |
Kontroldorne og tolerancedorne |
|
M&K(GEOM)-Taastrup |
Dimensional |
Ring gauges |
Length |
50
mm |
100
mm |
0,6
μm |
0,9
μm |
Længdemålemaskine og formtester (rundhed) |
120.44.08.02 |
Kontroldorne og tolerancedorne |
|
M&K(GEOM)-Taastrup |
Dimensional |
Ring gauges |
Length |
100
mm |
200
mm |
0,9
μm |
1,3
μm |
Længdemålemaskine og formtester (rundhed) |
120.44.08.02 |
Kontroldorne og tolerancedorne |
|
M&K(GEOM)-Taastrup |
Dimensional |
Ring gauges |
Length |
200
mm |
600
mm |
8
μm |
18
μm |
Længdemålemaskine og formtester (rundhed) |
120.44.08.02 |
Kontroldorne og tolerancedorne. Måleevne afhænger af dimension: (3+L x25)µm (L målt i meter) |
|
M&K(GEOM)-Taastrup |
Dimensional |
Plug gauges |
Length |
1
mm |
10
mm |
0,9
μm |
0,9
μm |
Længdemålemaskine og formtester (rundhed) |
120.44.08.03 |
Toleranceringe, referenceringe |
|
M&K(GEOM)-Taastrup |
Dimensional |
Plug gauges |
Length |
10
mm |
50
mm |
0,9
μm |
1,1
μm |
Længdemålemaskine og formtester (rundhed) |
120.44.08.03 |
Toleranceringe, referenceringe |
|
M&K(GEOM)-Taastrup |
Dimensional |
Plug gauges |
Length |
50
mm |
100
mm |
1,1
μm |
1,2
μm |
Længdemålemaskine og formtester (rundhed) |
120.44.08.03 |
Toleranceringe, referenceringe |
|
M&K(GEOM)-Taastrup |
Dimensional |
Plug gauges |
Length |
100
mm |
200
mm |
1,2
μm |
1,6
μm |
Længdemålemaskine og formtester (rundhed) |
120.44.08.03 |
Toleranceringe, referenceringe |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
0,5
mm |
10
mm |
0,06
μm |
0,06
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-510-550 |
Måleklodser: stål, hårdmetal og keramik. Klasse 0 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
10
mm |
25
mm |
0,06
μm |
0,07
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-510-550 |
Måleklodser: stål, hårdmetal og keramik. Klasse 0 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
25
mm |
50
mm |
0,07
μm |
0,08
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-510-550 |
Måleklodser: stål, hårdmetal og keramik. Klasse 0 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
50
mm |
75
mm |
0,08
μm |
0,1
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-510-550 |
Måleklodser: stål, hårdmetal og keramik. Klasse 0 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
75
mm |
100
mm |
0,1
μm |
0,12
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-510-550 |
Måleklodser: stål, hårdmetal og keramik. Klasse 0 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
0,5
mm |
10
mm |
0,07
μm |
0,07
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-511-551 |
Måleklodser: stål, hårdmetal og keramik. Klasse 1 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
10
mm |
25
mm |
0,07
μm |
0,08
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-511-551 |
Måleklodser: stål, hårdmetal og keramik. Klasse 1 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
25
mm |
50
mm |
0,08
μm |
0,1
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-511-551 |
Måleklodser: stål, hårdmetal og keramik. Klasse 1 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
50
mm |
75
mm |
0,1
μm |
0,12
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-511-551 |
Måleklodser: stål, hårdmetal og keramik. Klasse 1 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
75
mm |
100
mm |
0,12
μm |
0,14
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-511-551 |
Måleklodser: stål, hårdmetal og keramik. Klasse 1 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
0,5
mm |
10
mm |
0,08
μm |
0,08
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-512-552 |
Måleklodser: stål, hårdmetal og keramik. Klasse 2 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
10
mm |
25
mm |
0,08
μm |
0,1
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-512-552 |
Måleklodser: stål, hårdmetal og keramik. Klasse 2 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
25
mm |
50
mm |
0,1
μm |
0,12
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-512-552 |
Måleklodser: stål, hårdmetal og keramik. Klasse 2 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
50
mm |
75
mm |
0,12
μm |
0,15
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-512-552 |
Måleklodser: stål, hårdmetal og keramik. Klasse 2 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Gauge blocks |
Length |
75
mm |
100
mm |
0,15
μm |
0,18
μm |
Måleklodser |
DS/EN ISO 3650, intern procedure MK-512-552 |
Måleklodser: stål, hårdmetal og keramik. Klasse 2 |
|
M&K(GEOM)-Taastrup |
Dimensional |
Geometrical standards |
Length |
0
mm |
650
mm |
1,5
μm |
5
μm |
Universal målemikroskop |
200.49.16.02 |
Positionsmåling. Måleområde: 550x650 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Geometrical standards |
Length |
0
mm |
650
mm |
1,5
μm |
5
μm |
Universal målemikroskop |
200.49.16.02 |
Længdemåling. Måleområde: 550x650 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Geometrical standards |
Length |
0
mm |
650
mm |
2,5
μm |
5,6
μm |
Universal målemikroskop |
200.49.16.02 |
Rethedsmåling. Måleområde: 550x650 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Geometrical standards |
Length |
0
mm |
650
mm |
2,5
μm |
5,6
μm |
Universal målemikroskop |
200.49.16.02 |
Vinkelmåling. Måleområde: 550x650 mm |
|
M&K(GEOM)-Taastrup |
Dimensional |
Geometrical standards |
Length |
0
mm |
600
mm |
0,25
μm |
1,57
μm |
Vandret længdemålemaskine |
120.49.04.03 |
Længdemåling |
|
M&K(GEOM)-Taastrup |
Dimensional |
Geometrical standards |
Length |
10
mm |
50
mm |
0,6
μm |
0,6
μm |
Vandret længdemålemaskine |
120.44.08.09 |
Referencekugle, diameter |
|
M&K(GEOM)-Taastrup |
Dimensional |
Roundness standards |
Length |
0
mm |
300
mm |
0,08
μm |
0,08
μm |
FAG Formtester |
120.49.12.03 |
Rundheds-, retheds-, cylindricitets-, kast- og koaksialitetsmåling. Max. emnedimension: højde: 250 mm, diameter 300 mm og vægt 50 kg |
|
M&K(GEOM)-Taastrup |
Dimensional |
Coordinate Measuring Machines |
Length |
0
mm |
500
mm |
0,0015
mm |
0,0032
mm |
3 axes measuring machine |
200.44.17.01 |
Diameter and distance |
|
Taastrup / Aarhus |
Dimensional |
Coordinate Measuring Machines |
Length |
0
mm |
500
mm |
0,0032
mm |
0,0064
mm |
3 axes measuring machine |
200.44.17.01 |
Straightness, flatness, perpendicularity, angularity, parallelism, concentricity/coaxiality, circular run-out, total run-out. |
|
Taastrup / Aarhus |
Dimensional |
Inclinometer |
Degree (geometrical) (4
)
|
0
° |
90
° |
0,1
° |
0,1
° |
3 coordinate measuring machine |
200.44.06.01 |
For inclinometer with a resolution of 0,1° |
|
Tåstrup and Århus |
Dimensional |
Inclinometer |
Degree (geometrical) (3
)
|
0
° |
90
° |
0,05
° |
0,05
° |
3 coordinate measuring machine |
200.44.06.01 |
For inclinometer with a resolution of 0,05° |
|
Tåstrup and Århus |
Dimensional |
Inclinometer |
Degree (geometrical) (2
)
|
0
° |
90
° |
0,01
° |
0,01
° |
3 coordinate measuring machine |
200.44.06.01 |
For inclinometer with a resolution of 0,01° |
|
Tåstrup and Århus |
Dimensional |
Inclinometer |
Degree (geometrical) (1
)
|
0
° |
90
° |
0,003
° |
0,003
° |
3 coordinate measuring machine |
200.44.06.01 |
For inclinometer with a resolution of 0,001° |
|
Tåstrup and Århus |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
100
mΩ |
10
kΩ |
22
mΩ |
0,018
kΩ |
Installation testers and Insulation testers calibrator. |
Internal procedure Auto D1-11.1 |
2-Wire method
Resolution: 0.1 mΩ - 10 Ω
Continuity Test (Ω) and Earth Resistance (Ω) |
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
100
mΩ |
10
kΩ |
9
mΩ |
0,021
kΩ |
Installation and Insulation testers calibrator. |
Internal procedure Auto D1-11.1 |
4-Wire method
Resolution: 0.1 mΩ - 10 Ω
Continuity Test (Ω) and Earth Resistance (Ω) |
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
10
kΩ |
1000
MΩ |
0,017
kΩ |
4,5
MΩ |
Installation testers and Insulation testers calibrator. |
Internal procedure Auto D1-11.1 |
Insulation Resistance Test
Resolution: 1 Ω - 1 kΩ
Insulation Resistance (MΩ) |
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (DC) |
Resistance |
1
Ω |
1,8
kΩ |
0,094
Ω |
0,024
kΩ |
Installation testers and Insulation testers calibrator. |
Internal procedure Auto D1-11.1 |
Impedance Test:
Line Impedance Source (L-N), Loop Impedance Source (L-PE). |
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (DC) |
Voltage (DC) |
100
V |
1000
V |
2,9
V |
12
V |
Installation testers and Insulation testers calibrator. |
Internal procedure Auto D1-11.1 |
Insulation Test Voltage (V)
Resolution: 1V
|
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (AC) |
Voltage (AC) |
3
V |
600
V |
0,0066
V |
0,72
V |
Installation testers and Insulation testers calibrator. |
Internal procedure Auto D1-11.1 |
Resolution: 0,001 V - 0,1 V |
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (DC) |
Current (DC) |
0,5
mA |
5
mA |
0,015
mA |
0,088
mA |
Installation testers and Insulation testers calibrator. |
Internal procedure Auto D1-11.1 |
Insulation Test Current (mA)
Resolution: 0.1 mA
|
X |
AUTO-On site |
Electricity DC and LF |
Electrical equipment (AC) |
Current (AC) |
3
mA |
3
A |
0,027
mA |
0,027
A |
Installation testers and Insulation testers calibrator. |
Internal procedure Auto D1-11.1 |
RCD Trip Current
Resolution: 0.001 mA
|
X |
AUTO-On site |
|